MI-170意大利Milwaukee台式Nacl/Temp测试仪MI170
自动选择量程的EC/TDS/NaCl/Temp实验室检测仪
EC 和 TDS测量具有自动量程选择,自动设定最佳的分辨率。所有测试可在使用者设定的20度或25度下进行,温度自动补偿。可关闭自动温度补偿来测量实际电导率。四环电导电极准确完成电导率测试。GLP功能让使用者可以储存,调出系统数据。
可用USB和RS232与电脑连接
型号 |
MI170 | |
测量范围 |
EC |
0.00 to 29.99 μS/cm; 30.0 to 299.9 μS/cm; 300 to 2999 μS/cm; 3.00 to 29.99 mS/cm;30.0 to 200.0 mS/cm; up to 500.0 mS/cm |
TDS |
0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm);1.5 to | |
NACL |
0.0 to 400.0% | |
温度 |
-20.0 to | |
分辨率 |
EC |
0.01 μS/cm; 0.1 μS/cm; 1.0 μS/cm; 0.01 mS/cm; 0.1 mS/cm |
TDS |
0.01 mg/L; 0.1 mg/L; 1.0 mg/L; | |
NACL |
0.1% | |
温度 |
| |
精度 |
EC |
±1% of reading ±(0.05 μS/cm or 1 位) |
TDS |
±1% of reading ±(0.05 μS/cm or 1 位) | |
NACL |
±1% of reading | |
温度 |
±0.4°C / ±0.8°F | |
校准 |
EC |
1 点斜率,6个记忆校准点 (84.0 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) |
NACL |
1点, MA9066 | |
温度 |
2 点, 0 to | |
温度校准 |
自动补偿,-20.0 to 120.0°C / -4.0 to 248.0°F | |
温度系数 |
0.00 to 6.00%/°C | |
电极 |
MA814D/1 4环,内置温度传感器 | |
TDS系数 |
0.40 to 0.80 | |
LOG查询 |
每个范围50组历史数据 | |
GLP功能 |
最后的EC, NaCl 校准日期 | |
数据接口 |
RS232 / USB | |
适用环境 |
0 to | |
电池类型 |
12 VDC电源适配器 | |
尺寸/重量 |
230 x 160 x |
累计评价
0